Search the data

Instrumentation

CTD and auxiliary sensors

The CTD used on JC029 consisted of a 24-way stainless steel frame with the following instruments attached.

Instrument Serial Number Calibration date Comments
SBE 9 plus Underwater unit 09P-24680-0636 - -
SBE 3P Temperature sensor 03P-4151 2008-09-03 Primary sensor.
SBE 4C Conductivity sensor 04C-2571 2008-08-26 Primary sensor.
Digiquartz pressure sensor 83008 2008-09-10 -
SBE 3P Temperature sensor 03P-2919 2008-09-03 Secondary sensor, fin mounted.
SBE 4C Conductivity sensor 04C-2450 2008-08-26 Secondary sensor, fin mounted.
SBE 5T Submersible pump 05T-4166 2008-08-26 Primary sensor.
SBE 5T Submersible pump 05T-2793 2008-08-26 Secondary senor, fin mounted.
SBE 43 Oxygen 43-0363 2008-09-09 -
Chelsea MKIII Aquatracka Fluorometer 088108 2008-01-09 -
Benthos PSA-916T Altimeter 1040 - -
Chelsea MKII Alphatracka 25 cm path Transmissometer 161045 2005-09-08 -
Wetlabs BBRTD backscatter 115R 2008-05-13 -
Teledyne RDI 300 kHz Workhorse Monitor lowered ADCP 9192 - Downward-looking master configuration, failed on beam 3 during CTD025.
Teledyne RDI 300 kHz Workhorse Monitor lowered ADCP 5415 - Upward-looking slave configuration, failed on beam 2 during CTD010.
Teledyne RDI 300 kHz Workhorse Monitor lowered ADCP 9191 - Upward-looking slave configuration from casts CTD012 to CTD014. Used as downward-looking master from cast CTD026 onwards.
Teledyne RDI Broadband 150kHz lowered ADCP 1503 - Downward-looking slave configuration, removed from casts CTD016 to CTD040.

A complete spare CTD suite including instruments was available for use but was not required.