Search the data

Metadata Report for BODC Series Reference Number 1222437


Metadata Summary

Data Description

Data Category Meteorology -unspecified
Instrument Type
NameCategories
TriOS RAMSES ACC-UV Hyperspectral UVA and UVB Radiometer  radiometers
Instrument Mounting research vessel
Originating Country United Kingdom
Originator Dr Tim Smyth
Originating Organization Plymouth Marine Laboratory
Processing Status banked
Online delivery of data Download available - Ocean Data View (ODV) format
Project(s) -
 

Data Identifiers

Originator's Identifier IUV_LATLON_121117
BODC Series Reference 1222437
 

Time Co-ordinates(UT)

Start Time (yyyy-mm-dd hh:mm) 2012-11-17 06:35
End Time (yyyy-mm-dd hh:mm) 2012-11-17 22:10
Nominal Cycle Interval 300.0 seconds
 

Spatial Co-ordinates

Start Latitude 40.16880 S ( 40° 10.1' S )
End Latitude 41.58190 S ( 41° 34.9' S )
Start Longitude 42.48850 W ( 42° 29.3' W )
End Longitude 44.71910 W ( 44° 43.1' W )
Positional Uncertainty 0.0 to 0.01 n.miles
Minimum Sensor or Sampling Depth -15.0 m
Maximum Sensor or Sampling Depth -15.0 m
Minimum Sensor or Sampling Height -
Maximum Sensor or Sampling Height -
Sea Floor Depth -
Sea Floor Depth Source -
Sensor or Sampling Distribution Fixed common depth - All sensors are grouped effectively at the same depth which is effectively fixed for the duration of the series
Sensor or Sampling Depth Datum Approximate - Depth is only approximate
Sea Floor Depth Datum -
 

Parameters

BODC CODERankUnitsTitle
AADYAA011DaysDate (time from 00:00 01/01/1760 to 00:00 UT on day)
AAFDZZ011DaysTime (time between 00:00 UT and timestamp)
ACYCAA011DimensionlessSequence number
ALATGP011DegreesLatitude north relative to WGS84 by unspecified GPS system
ALONGP011DegreesLongitude east relative to WGS84 by unspecified GPS system
VSRAUVAW1Watts per square metreDownwelling vector irradiance as energy of electromagnetic radiation (ultra-violet A wavelengths) in the atmosphere by cosine-collector radiometer
VSRAUVBW1Watts per square metreDownwelling vector irradiance as energy of electromagnetic radiation (ultra-violet B wavelengths) in the atmosphere by cosine-collector radiometer

Definition of Rank

  • Rank 1 is a one-dimensional parameter
  • Rank 2 is a two-dimensional parameter
  • Rank 0 is a one-dimensional parameter describing the second dimension of a two-dimensional parameter (e.g. bin depths for moored ADCP data)

Problem Reports

No Problem Report Found in the Database


Data Access Policy

Public domain data

These data have no specific confidentiality restrictions for users. However, users must acknowledge data sources as it is not ethical to publish data without proper attribution. Any publication or other output resulting from usage of the data should include an acknowledgment.

The recommended acknowledgment is

"This study uses data from the data source/organisation/programme, provided by the British Oceanographic Data Centre and funded by the funding body."


Narrative Documents

TriOS RAMSES-ACC-UV radiometer

Basic Information

This is a spectral radiometer measuring over the UV band range (280-500 nm) with a cosine-response to measure irradiance. It is fitted with a 256 channel silicon photodiode array with channels giving 2.2 nm/pixel and a spectral accuracy of 02 nm. Typical saturation (4ms integration time) is 20 W m-2 nm-1 sr-1 (at 300 nm), 17 W m-2 nm-1 sr -1 (at 360nm), 18W m-2 nm-1 sr -1 (at 500nm). Typical NEI (8ms integration time) is 0.85 µW m-2 nm-1 sr -1 (at 300nm), 0.75 µW m-2 nm-1 sr -1 (at 360nm), 0.80 µW m-2 nm-1 sr -1 (at 500nm). Acuuracy is quoted as being better than 6-10 per cent depending on spectral range.

Performance Specifications

Optics

wavelength range* 280 - 500 nm
detector type* 256 channel silicon photodiode array
spectral sampling*: 2.2 nm/pixel
spectral accuracy: 0.2 nm
usable channels: 100
typical saturation (4ms integration time): 20 W m-2 nm-1 sr-1 (at 300nm)
17 W m-2 nm-1 sr-1 (at 360nm)
18 W m-2 nm-1 sr-1 (at 500nm)
typical NEI (8ms integration time): 0.85 µW m-2 nm-1 sr-1 (at 300nm)
0.75 µW m-2 nm-1 sr-1 (at 360nm)
0.80 µW m-2 nm-1 sr-1 (at 500nm)

Detection

collector type: cosine response
accuracy: better than 6-10% (depending on spectral range)

Electrical

integration time: 4 ms - 8 sec (user selectable or auto [4 ms - 4096 ms])
telemetry data interface: RS232 or Serial Bus
data rate (RS232): 1,200 - 19,200 baud
power requirements: 1.5 - 11 VDC
0.85mW (data acquisition active)
100mW (interface active)
0.5mW (stand-by modus)
connector: SubConn micro 5 pin male

Physical

size: dia. 4.83 cm x 26 cm (without connector)
weight in air: <1.0 kg (stainless steel / POM housing)
depth range: 300m
operating temperature: -10°C to +50°C

* specifications from Carl ZEISS, Germany

Further information may be found in the TriOS RAMSES brochure.

AMT22 RRS James Cook (JC079) Surface UV Instrumentation

Instrument Serial number Parameter
Trios Rameses ACC UV Not supplied UVA and UVB

AMT22 RRS James Cook (JC079) Surface UV Measurements Processing Document

Originator's Processing

A Trios Rameses ACC UV sensor was set up high on the ship's starboard winch control cabin and configured to log hyperspectral UV between 200 and 500 nm at 2.5nm resolutuion every 5 minutes through daylight hours. The data can either be kept at hyperspectral or integrated over broadband (UV-A and UV-B) ranges.

The data were fully calibarated by a calibration carried out by TRIOS in May 2011. The data were split into the UVA and UVB components using the idl rouitine trios_uv_deck.pro where UVA and UVB were defined as 320< λ <380 nm and 280< λ <320 nm respectively. The data were subsequently placed into daily files.

Investigations have found that the calibration of the instruments drifts dramatically during cruises. The data have not been corrected for this drift.

Data were submitted to BODC as text files with each individual file name dusing the convention yymmdd_uva_uvb.txt as each file represented a a day of UV measurement on the cruise.

BODC Processing

The data were received and archived to the BODC system. Due to the lack of longitude and latitude in the originator's files the underway latitude and longitude in the BODC bank were concatenated with the UV files. The data were then transferred to the BODC format and loaded to the databse. databse.

Parameters

Parameter codes defined in the BODC parameter dictionary were assigned to the variables as shown in the table below:

Originator's Parameter Originator's Units BODC Parameter Code BODC Parameter Units BODC Description
UVA W/(m^2) VSRAUVAW Watts per square metre Downwelling vector irradiance as energy (ultra-violet A wavelengths) in the atmosphere by cosine-collector radiometer
UVB W/(m^2) VSRAUVBW Watts per square metre Downwelling vector irradiance as energy (ultra-violet B wavelengths) in the atmosphere by cosine-collector radiometer
N/A N/A ALATGP01 Degrees Latitude north (WGS84) by unspecified GPS system
N/A N/A ALONGP01 Degrees Longitude east (WGS84) by unspecified GPS system

Screening

All the data on 10/10/2012 were flagged 'M' as the values were considerably lower than the data on subsequent dates at the equivalent times. In addition to this no diurnal signal could be seen.

The rest of the data appears to have a strong diurnal signal with a few spikes which have been left unflagged as they are considered to be within the realms of possibility.


Project Information


No Project Information held for the Series

Data Activity or Cruise Information

Cruise

Cruise Name JC079 (AMT22)
Departure Date 2012-10-10
Arrival Date 2012-11-24
Principal Scientist(s)Glen A Tarran (Plymouth Marine Laboratory)
Ship RRS James Cook

Complete Cruise Metadata Report is available here


Fixed Station Information


No Fixed Station Information held for the Series


BODC Quality Control Flags

The following single character qualifying flags may be associated with one or more individual parameters with a data cycle:

Flag Description
Blank Unqualified
< Below detection limit
> In excess of quoted value
A Taxonomic flag for affinis (aff.)
B Beginning of CTD Down/Up Cast
C Taxonomic flag for confer (cf.)
D Thermometric depth
E End of CTD Down/Up Cast
G Non-taxonomic biological characteristic uncertainty
H Extrapolated value
I Taxonomic flag for single species (sp.)
K Improbable value - unknown quality control source
L Improbable value - originator's quality control
M Improbable value - BODC quality control
N Null value
O Improbable value - user quality control
P Trace/calm
Q Indeterminate
R Replacement value
S Estimated value
T Interpolated value
U Uncalibrated
W Control value
X Excessive difference

SeaDataNet Quality Control Flags

The following single character qualifying flags may be associated with one or more individual parameters with a data cycle:

Flag Description
0 no quality control
1 good value
2 probably good value
3 probably bad value
4 bad value
5 changed value
6 value below detection
7 value in excess
8 interpolated value
9 missing value
A value phenomenon uncertain
B nominal value
Q value below limit of quantification