Search the data

AMT14 Bench salinometer measurements from CTD rosette bottle samples

Originator's Protocol for Data Acquisition and Analysis

Samples were taken from CTD rossette bottles to provide a calibration for the SeaBird conductivity sensors. During the cruise 78 samples were collected and processed for salinity measurement.

A Guildline Autosal 8400B (BAS) salinometer was available for use on the cruise. The main salinometer was located in the Prep Laboratory and operated at 24°C bath temperature in 21-24°C ambient lab temperature. No problems with the salinometer or its performance were noted. All samples were processed according to WOCE standards and protocols.

References Cited

BODC Data Processing Procedures

Data were submitted with the Post-Cruise Archive as measurements listed against sample bottles and archived in accession SOC040057. The samples were matched to metadata (cast and depth) using the sample numbers that were recorded on the CTD bottle firing logsheets and bench salinometer logsheets provided to BODC. There were no discrepancies compared to metadata already loaded to the database.

No unit conversions were necessary as the units the data were supplied in the files matched the units for the parameter codes in the BODC Parameter Dictionary. The data were loaded to the database using established BODC procedures.

A parameter mapping table is provided below;

Originator's Parameter Units Description BODC Parameter Code Units Comments
Salinity (mean) Dimensionless Practical salinity of the water body by bench salinometer and computation using UNESCO 1983 algorithm PSALBSTX Dimensionless n/a

Data Quality Report

The originator did not advise BODC of any data quality issues.

Problem Report

Not relevant to this data set.