Aanderaa Recording Current Meter Model 7/8

Manufacturer's specifications: recording unit height 49.5cm (RCM8 52.0cm), diameter 12.8cm, vane size 48.5x50.0cm. Meter is designed for depths down to 2000m (RCM8 6000m). It incorporates a spindle which is shackled to the mooring line. The meter is attached to the spindle through a gimbal mounting which permits a maximum 27° deviation of the spindle from the vertical, the meter still remaining horizontal.

Meter comprises :

  1. Paddle wheel rotor magnetically coupled to an electronic counter

  2. Vane, which aligns instrument with current flow, has a balance weight ensuring static balance and tail fins to ensure dynamic balance in flows up to 250cm/s.

  3. Magnetic compass (needle is clamped to potentiometer ring) - direction recorded with 0.35° resolution, 5° accuracy for speeds 5 to 100cm/s, 7.5° accuracy for remaining speeds within 2.5 to 200cm/s range.

  4. Quartz clock, accuracy better than 2 sec/day within temperature range 0 to 20°C.

  5. Thermistor (temperature sensor), standard range -2.46 to 21.48°C (max on high range 36.04°C), accuracy 0.05°C, resolution 0.1 per cent of range, 63 per cent response time 12sec.

  6. Inductive cell conductivity sensor (optional), range 0 to 70mmho/cm standard resolution 0.1 per cent of range.

  7. Silicon piezoresistive bridge, standard range 0 to 3000 psi (RCM8 to 9000 psi), resolution 0.1% of range.

  8. Self balancing potentiometer which converts the output from each sensor into a 10 bit binary number for storage on magnetic tape.

  9. Associated electronics.

A built-in clock triggers the instrument at preset intervals and up to six channels are sampled in sequence. Channel 1 is a fixed reference reading for control purposes and data identification. Channels 2, 3 and 4 represent measurement of temperature, conductivity and pressure. Channels 5 and 6 represent the VECTOR AVERAGED current speed and direction since the previous triggering of the instrument. The number of rotor revolutions and the direction is sampled every 12 seconds and broken into North and East components. Successive components are added and recorded as speed and direction. For recording intervals longer than 10 minutes, speed and direction are sampled 1/50th of recording interval.

It has become common practice in some laboratories to deploy these meters as temperature and conductivity loggers without current measuring capabilities.

The following link will provide the manufacturer specifications:

Manufacturer specifications