Metadata Report for BODC Series Reference Number 506836


Metadata Summary

Data Description

Data Category CTD or STD cast
Instrument Type
NameCategories
Neil Brown MK3 CTD  CTD; water temperature sensor; salinity sensor; dissolved gas sensors
SeaTech transmissometer  transmissometers
Chelsea Technologies Group Aquatracka fluorometer  fluorometers
Instrument Mounting research vessel
Originating Country United Kingdom
Originator -
Originating Organization Plymouth Marine Laboratory
Processing Status QC in progress
Project(s) -
 

Data Identifiers

Originator's Identifier ctd001.d
BODC Series Reference 506836
 

Time Co-ordinates(UT)

Start Time (yyyy-mm-dd hh:mm) 1992-03-04 23:43
End Time (yyyy-mm-dd hh:mm) -
Nominal Cycle Interval 1.0 seconds
 

Spatial Co-ordinates

Latitude 37.00480 N ( 37° 0.3' N )
Longitude 27.04980 W ( 27° 3.0' W )
Positional Uncertainty Unspecified
Minimum Sensor Depth 2.48 m
Maximum Sensor Depth 2482.25 m
Minimum Sensor Height -138.25 m
Maximum Sensor Height 2341.52 m
Sea Floor Depth 2344.0 m
Sensor Distribution Variable common depth - All sensors are grouped effectively at the same depth, but this depth varies significantly during the series
Sensor Depth Datum Instantaneous - Depth measured below water line or instantaneous water body surface
Sea Floor Depth Datum Instantaneous - Depth measured below water line or instantaneous water body surface
 

Parameters

BODC CODE Rank Units Short Title Title
AADYAA01 1 Days Date(Loch_Day) Date (time from 00:00 01/01/1760 to 00:00 UT on day)
AAFDZZ01 1 Days Time(Day_Fract) Time (time between 00:00 UT and timestamp)
ACYCAA01 1 Dimensionless Record_No Sequence number
ATTNMR01 1 per metre Atten_red Attenuation (red light wavelength) per unit length of the water body by 20 or 25cm path length transmissometer
DOXYPR01 1 Micromoles per litre WC_dissO2_Beck Concentration of oxygen {O2 CAS 7782-44-7} per unit volume of the water body [dissolved plus reactive particulate phase] by in-situ Beckmann probe
FVLTAQ01 1 Volts AqVolt Instrument output (voltage) by in-situ Aquatracka chlorophyll fluorometer
PRESPR01 1 Decibars Pres_Z Pressure (spatial co-ordinate) exerted by the water body by profiling pressure sensor and corrected to read zero at sea level
PSALPR01 1 Dimensionless P_sal Practical salinity of the water body by conductivity cell and computation using UNESCO 1983 algorithm
TEMPST01 1 Degrees Celsius WC_temp_CTD Temperature of the water body by CTD or STD
 

Definition of Rank

  • Rank 1 is a one-dimensional parameter
  • Rank 2 is a two-dimensional parameter
  • Rank 0 is a one-dimensional parameter describing the second dimension of a two-dimensional parameter (e.g. bin depths for moored ADCP data)

Problem Reports

No Problem Report Found in the Database


Data Access Policy

Open Data supplied by Natural Environment Research Council (NERC)

You must always use the following attribution statement to acknowledge the source of the information: "Contains data supplied by Natural Environment Research Council."


Narrative Documents

Neil Brown MK3 CTD

The Neil Brown MK3 conductivity-temperature-depth (CTD) profiler consists of an integral unit containing pressure, temperature and conductivity sensors with an optional dissolved oxygen sensor in a pressure-hardened casing. The most widely used variant in the 1980s and 1990s was the MK3B. An upgrade to this, the MK3C, was developed to meet the requirements of the WOCE project.

The MK3C includes a low hysteresis, titanium strain gauge pressure transducer. The transducer temperature is measured separately, allowing correction for the effects of temperature on pressure measurements. The MK3C conductivity cell features a free flow, internal field design that eliminates ducted pumping and is not affected by external metallic objects such as guard cages and external sensors.

Additional optional sensors include pH and a pressure-temperature fluorometer. The instrument is no longer in production, but is supported (repair and calibration) by General Oceanics.

Specifications

These specification apply to the MK3C version.

Pressure Temperature Conductivity
Range

6500 m

3200 m (optional)

-3 to 32°C 1 to 6.5 S cm -1
Accuracy

0.0015% FS

0.03% FS < 1 msec

0.0005°C

0.003°C < 30 msec

0.0001 S cm -1

0.0003 S cm -1 < 30 msec

Further details can be found in the specification sheet .

Aquatracka fluorometer

The Chelsea Instruments Aquatracka is a logarithmic response fluorometer. It uses a pulsed (5.5 Hz) xenon light source discharging between 320 and 800 nm through a blue filter with a peak transmission of 420 nm and a bandwidth at half maximum of 100 nm. A red filter with sharp cut off, 10% transmission at 664 nm and 678 nm, is used to pass chlorophyll-a fluorescence to the sample photodiode.

The instrument may be deployed either in a through-flow tank, on a CTD frame or moored with a data logging package.

Further details can be found in the manufacturer's specification sheet .

SeaTech Transmissometer

Introduction

The transmissometer is designed to accurately measure the the amount of light transmitted by a modulated Light Emitting Diode (LED) through a fixed-length in-situ water column to a synchronous detector.

Specifications

Notes

The instrument can be interfaced to Aanderaa RCM7 current meters. This is achieved by fitting the transmissometer in a slot cut into a customized RCM4-type vane.

A red LED (660 nm) is used for general applications looking at water column sediment load. However, green or blue LEDs can be fitted for specilised optics applications. The light source used is identified by the BODC parameter code.

Further details can be found in the manufacturer's Manual .

RRS Charles Darwin 66 CTD Data Documentation

Document text in preparation.


Project Information


No Project Information held for the Series

Data Activity or Cruise Information

Cruise

Cruise Name CD66
Departure Date 1992-03-04
Arrival Date 1992-04-06
Principal Scientist(s)Robin D Pingree (Plymouth Marine Laboratory)
Ship RRS Charles Darwin

Complete Cruise Metadata Report is available here


Fixed Station Information


No Fixed Station Information held for the Series


BODC Quality Control Flags

The following single character qualifying flags may be associated with one or more individual parameters with a data cycle:

Flag Description
Blank Unqualified
< Below detection limit
> In excess of quoted value
A Taxonomic flag for affinis (aff.)
B Beginning of CTD Down/Up Cast
C Taxonomic flag for confer (cf.)
D Thermometric depth
E End of CTD Down/Up Cast
G Non-taxonomic biological characteristic uncertainty
H Extrapolated value
I Taxonomic flag for single species (sp.)
K Improbable value - unknown quality control source
L Improbable value - originator's quality control
M Improbable value - BODC quality control
N Null value
O Improbable value - user quality control
P Trace/calm
Q Indeterminate
R Replacement value
S Estimated value
T Interpolated value
U Uncalibrated
W Control value
X Excessive difference

SeaDataNet Quality Control Flags

The following single character qualifying flags may be associated with one or more individual parameters with a data cycle:

Flag Description
0 no quality control
1 good value
2 probably good value
3 probably bad value
4 bad value
5 changed value
6 value below detection
7 value in excess
8 interpolated value
9 missing value
A value phenomenon uncertain