Search the data

Metadata Report for BODC Series Reference Number 914644


Metadata Summary

Data Description

Data Category CTD or STD cast
Instrument Type
NameCategories
Sea-Bird SBE 25 Sealogger CTD  CTD; water temperature sensor; salinity sensor
Instrument Mounting research vessel
Originating Country France
Originator Dr Jean-Marie Jouanneau
Originating Organization University of Bordeaux 1, Environnements et Paléoenvironnements Océaniques et Continentaux
Processing Status banked
Online delivery of data Download available - Ocean Data View (ODV) format
Project(s) OMEX I
 

Data Identifiers

Originator's Identifier CTD01
BODC Series Reference 914644
 

Time Co-ordinates(UT)

Start Time (yyyy-mm-dd hh:mm) 1993-09-03 08:22
End Time (yyyy-mm-dd hh:mm) -
Nominal Cycle Interval 2.0 decibars
 

Spatial Co-ordinates

Latitude 47.73000 N ( 47° 43.8' N )
Longitude 5.38217 W ( 5° 22.9' W )
Positional Uncertainty 0.0 to 0.01 n.miles
Minimum Sensor or Sampling Depth 4.96 m
Maximum Sensor or Sampling Depth 121.93 m
Minimum Sensor or Sampling Height 7.06 m
Maximum Sensor or Sampling Height 124.04 m
Sea Floor Depth 129.0 m
Sea Floor Depth Source DATAHEAD
Sensor or Sampling Distribution Variable common depth - All sensors are grouped effectively at the same depth, but this depth varies significantly during the series
Sensor or Sampling Depth Datum Instantaneous - Depth measured below water line or instantaneous water body surface
Sea Floor Depth Datum Instantaneous - Depth measured below water line or instantaneous water body surface
 

Parameters

BODC CODERankUnitsTitle
CPHLPR011Milligrams per cubic metreConcentration of chlorophyll-a {chl-a CAS 479-61-8} per unit volume of the water body [particulate >unknown phase] by in-situ chlorophyll fluorometer
PRESPR011DecibarsPressure (spatial coordinate) exerted by the water body by profiling pressure sensor and correction to read zero at sea level
TEMPST011Degrees CelsiusTemperature of the water body by CTD or STD
TURBPR011Nephelometric Turbidity UnitsTurbidity of water in the water body by in-situ optical backscatter measurement and laboratory calibration against formazin

Definition of Rank

  • Rank 1 is a one-dimensional parameter
  • Rank 2 is a two-dimensional parameter
  • Rank 0 is a one-dimensional parameter describing the second dimension of a two-dimensional parameter (e.g. bin depths for moored ADCP data)

Problem Reports

No Problem Report Found in the Database


Data Access Policy

Public domain data

These data have no specific confidentiality restrictions for users. However, users must acknowledge data sources as it is not ethical to publish data without proper attribution. Any publication or other output resulting from usage of the data should include an acknowledgment.

The recommended acknowledgment is

"This study uses data from the data source/organisation/programme, provided by the British Oceanographic Data Centre and funded by the funding body."


Narrative Documents

Sea-Bird SBE 25 SEALOGGER CTD

The SBE 25 SEALOGGER is a research-quality CTD profiling system used for coastal, estuarine and, can also be a practical option, for deep-water work. It is easily configured in the field for a wide range of auxiliary sensors. The SEALOGGER is self-powered, requires no conductive cable, and is designed for use up to 6800 meters (10,000 psia). It uses the SBE3F temperature and SBE4 conductivity sensors as well as having an external strain gauge pressure sensor. It provides pump-controlled T-C ducted flow, samples at 8 Hz, records internally and provides simultaneous real-time data via its built-in RS-232 interface.

The standard CTD unit comes with a plastic housing (rated to 600 m), although this can be replaced by aluminium housing for depths up to 6800 m.

Specifications

Parameter SBE 25
Temperature

Range: -5 to +35 °C

Accuracy: 0.002 °C

Resolution: 0.0003 °C

Conductivity

Range: 0 to 7 S m-1 (0 to 70 mmho cm-1)

Accuracy: 0.0003 S m-1

Resolution: 0.00004 S m-1

Strain gauge pressure sensor

Range: 0 to 20, 100, 350, 600, 1000, 2000, 3500, 7000 metres (expressed in metres of deployment depth capability)

Accuracy: 0.1% of full scale range

Resolution: 0.015% of full scale range

Options and accessories

Additional sensors can be attached to the CTD, including:

  • Dissolved Oxygen (SBE 43 DO Sensor)
  • pH (SBE 18 pH Sensor or SBE 27 pH/ORP Sensor)
  • fluorescence
  • radiance (PAR)
  • light transmission
  • optical backscatter (turbidity)

The SBE 5T titanium pump can be used in place of SBE 5P pump. Further details can be found in the manufacturer's SBE 25 instrument specification.

RV Côte D'Aquitaine

Instrumentation and Shipboard Protocols

NAOMEX 1 (BODC mnemonic NAOX1) was a University of Bordeaux cruise on the research vessel Côte d'Acquitaine. The CTD profiles were taken with the SeaBird SBE25 system fitted with a Chelsea Instruments fluorometer, a nephelometer and an oxygen membrane of the Beckman (non-pulsed) type.

Data Acquisition

The SBE25 is normally a self-logging instrument with the data downloaded onto a PC running the SeaBird DATCNV software. The raw data files are converted into ASCII files with the data in oceanographic units on the basis of coefficients held in a calibration file.

Post-Cruise Processing

ASCII DATCNV output files were supplied to BODC.

Reformatting

The data as supplied had been binned to 0.25 db with temperature, practical salinity, chlorophyll (nominal units), oxygen (ml/l) and optical backscatter (nominal units).

The data were converted into the BODC internal format (PXF) to allow the use of in-house software tools, notably the workstation graphics editor. In addition to reformatting, the transfer program applied the following modifications to the data:

Dissolved oxygen was converted from ml/l to µM by multiplying the values by 44.66.

The chlorophyll was converted back to a voltage by a natural log transform to conform to the requirements of the BODC CTD data handling system. On retrieval, the data as supplied are reproduced.

Editing and Quality Control

Using a custom in-house graphics editor, the limits of the downcast were manually flagged and any obvious spikes identified were manually flagged 'suspect'.

Once screened, the CTD downcasts (between the flagged limits) were loaded into a database under the Oracle relational database management system.

The manual inspection procedures revealed that there were serious problems with the salinity data. The data were plagued by frequent oscillations with an amplitude of up to 0.5 PSU. This, combined with the fact that the calibration accuracy of the salinity is totally unknown, led to the decision to delete the entire salinity channel from the data set.

The form of the dissolved oxygen profiles, with the exception of the upper 50m of the first cast, looked perfectly reasonable. However, the absolute values were obviously seriously low with saturations of below 80% at the surface. No water bottle data were available with which to correct the CTD oxygen data. Consequently, this channel was also deleted.

Calibrations

No sample data were available to calibrate any of the channels. All data are therefore the result of manufacturer's calibrations of unknown date and no guarantee can be given as to the accuracy of the data.

Data Reduction

Once all screening and calibration procedures were completed, the data set was binned to 2 db (casts deeper than 100 db) or 1 db (casts shallower than 100 db). Downcast values corresponding to the bottle firing depths were incorporated into the database.

Data Warnings

The salinity data were subject to severe oscillations and have been deleted.

The absolute dissolved oxygen data were obviously wrong and no sample data were available for their calibration. Consequently, the oxygen data have also been deleted.

No independent checks, such as reversing thermometer data, were available for the verification of the temperature data. The accuracy of this channel is therefore unknown.

The fluorometer has not been calibrated against extracted chlorophyll data. The absolute values may therefore be meaningless.

The nephelometer data are in arbitrary units and their absolute values have no meaning.


Project Information

Ocean Margin EXchange (OMEX) I

Introduction

OMEX was a European multidisciplinary oceanographic research project that studied and quantified the exchange processes of carbon and associated elements between the continental shelf of western Europe and the open Atlantic Ocean. The project ran in two phases known as OMEX I (1993-1996) and OMEX II - II (1997-2000), with a bridging phase OMEX II - I (1996-1997). The project was supported by the European Union under the second and third phases of its MArine Science and Technology Programme (MAST) through contracts MAS2-CT93-0069 and MAS3-CT97-0076. It was led by Professor Roland Wollast from Université Libre de Bruxelles, Belgium and involved more than 100 scientists from 10 European countries.

Scientific Objectives

The aim of the Ocean Margin EXchange (OMEX) project was to gain a better understanding of the physical, chemical and biological processes occurring at the ocean margins in order to quantify fluxes of energy and matter (carbon, nutrients and other trace elements) across this boundary. The research culminated in the development of quantitative budgets for the areas studied using an approach based on both field measurements and modeling.

OMEX I (1993-1996)

The first phase of OMEX was divided into sub-projects by discipline:

  • Physics
  • Biogeochemical Cycles
  • Biological Processes
  • Benthic Processes
  • Carbon Cycling and Biogases

This emphasises the multidisciplinary nature of the research.

The project fieldwork focussed on the region of the European Margin adjacent to the Goban Spur (off the coast of Brittany) and the shelf break off Tromsø, Norway. However, there was also data collected off the Iberian Margin and to the west of Ireland. In all a total of 57 research cruises (excluding 295 Continuous Plankton Recorder tows) were involved in the collection of OMEX I data.

Data Availability

Field data collected during OMEX I have been published by BODC as a CD-ROM product, entitled:

  • OMEX I Project Data Set (two discs)

Further descriptions of this product and order forms may be found on the BODC web site.

The data are also held in BODC's databases and subsets may be obtained by request from BODC.


Data Activity or Cruise Information

Cruise

Cruise Name Naomex 1
Departure Date 1993-09-02
Arrival Date 1993-09-11
Principal Scientist(s)Jean-Marie Jouanneau (University of Bordeaux 1, Environnements et Paléoenvironnements Océaniques et Continentaux)
Ship Cote d'Aquitaine

Complete Cruise Metadata Report is available here


Fixed Station Information


No Fixed Station Information held for the Series


BODC Quality Control Flags

The following single character qualifying flags may be associated with one or more individual parameters with a data cycle:

Flag Description
Blank Unqualified
< Below detection limit
> In excess of quoted value
A Taxonomic flag for affinis (aff.)
B Beginning of CTD Down/Up Cast
C Taxonomic flag for confer (cf.)
D Thermometric depth
E End of CTD Down/Up Cast
G Non-taxonomic biological characteristic uncertainty
H Extrapolated value
I Taxonomic flag for single species (sp.)
K Improbable value - unknown quality control source
L Improbable value - originator's quality control
M Improbable value - BODC quality control
N Null value
O Improbable value - user quality control
P Trace/calm
Q Indeterminate
R Replacement value
S Estimated value
T Interpolated value
U Uncalibrated
W Control value
X Excessive difference

SeaDataNet Quality Control Flags

The following single character qualifying flags may be associated with one or more individual parameters with a data cycle:

Flag Description
0 no quality control
1 good value
2 probably good value
3 probably bad value
4 bad value
5 changed value
6 value below detection
7 value in excess
8 interpolated value
9 missing value
A value phenomenon uncertain
B nominal value
Q value below limit of quantification