Search the data

Metadata Report for BODC Series Reference Number 1077320


Metadata Summary

Data Description

Data Category CTD or STD cast
Instrument Type
NameCategories
Sea-Bird SBE 43 Dissolved Oxygen Sensor  dissolved gas sensors
Sea-Bird SBE 911plus CTD  CTD; water temperature sensor; salinity sensor
Tritech PA-200 Altimeter  altimeters
WET Labs {Sea-Bird WETLabs} C-Star transmissometer  transmissometers
Biospherical QCD-905L underwater PAR sensor  radiometers
Chelsea Technologies Group Aquatracka III fluorometer  fluorometers
Instrument Mounting research vessel
Originating Country United Kingdom
Originator Dr Graham Quartly
Originating Organization National Oceanography Centre, Southampton
Processing Status banked
Online delivery of data Download available - Ocean Data View (ODV) format
Project(s) Oceans 2025
Oceans 2025 Theme 10
Oceans 2025 Theme 10 SO6
 

Data Identifiers

Originator's Identifier CTD_JR194_001_2DB
BODC Series Reference 1077320
 

Time Co-ordinates(UT)

Start Time (yyyy-mm-dd hh:mm) 2008-12-12 20:30
End Time (yyyy-mm-dd hh:mm) 2008-12-12 20:34
Nominal Cycle Interval 2.0 decibars
 

Spatial Co-ordinates

Latitude 51.77160 S ( 51° 46.3' S )
Longitude 57.58210 W ( 57° 34.9' W )
Positional Uncertainty 0.0 to 0.01 n.miles
Minimum Sensor or Sampling Depth 2.97 m
Maximum Sensor or Sampling Depth 102.08 m
Minimum Sensor or Sampling Height 14.92 m
Maximum Sensor or Sampling Height 114.03 m
Sea Floor Depth 117.0 m
Sea Floor Depth Source DATAHEAD
Sensor or Sampling Distribution Variable common depth - All sensors are grouped effectively at the same depth, but this depth varies significantly during the series
Sensor or Sampling Depth Datum Instantaneous - Depth measured below water line or instantaneous water body surface
Sea Floor Depth Datum Instantaneous - Depth measured below water line or instantaneous water body surface
 

Parameters

BODC CODERankUnitsTitle
ACYCAA011DimensionlessSequence number
AHSFZZ011MetresHeight (spatial coordinate) relative to bed surface in the water body
CNCLCCI11Siemens per metreElectrical conductivity of the water body by in-situ conductivity cell and calibration against independent measurements
CPHLPM011Milligrams per cubic metreConcentration of chlorophyll-a {chl-a CAS 479-61-8} per unit volume of the water body [particulate >unknown phase] by in-situ chlorophyll fluorometer and manufacturer's calibration applied
DEPHPR011MetresDepth (spatial coordinate) relative to water surface in the water body by profiling pressure sensor and conversion to seawater depth using UNESCO algorithm
DOXYSU011Micromoles per litreConcentration of oxygen {O2 CAS 7782-44-7} per unit volume of the water body [dissolved plus reactive particulate phase] by Sea-Bird SBE 43 sensor and no calibration against sample data
IRRDSD011MicroEinsteins per square metre per secondDownwelling vector irradiance as photons standard deviation of electromagnetic radiation (PAR wavelengths) in the water body by cosine-collector radiometer
OXYSZZ011PercentSaturation of oxygen {O2 CAS 7782-44-7} in the water body [dissolved plus reactive particulate phase]
POPTPZ011PercentTransmittance (unspecified wavelength) per unit length of the water body by transmissometer and correction to a path length of 1m
PRESPR011DecibarsPressure (spatial coordinate) exerted by the water body by profiling pressure sensor and correction to read zero at sea level
PSALCC011DimensionlessPractical salinity of the water body by CTD and computation using UNESCO 1983 algorithm and calibration against independent measurements
SIGTPR011Kilograms per cubic metreSigma-theta of the water body by CTD and computation from salinity and potential temperature using UNESCO algorithm
TEMPCU011Degrees CelsiusTemperature of the water body by CTD and NO verification against independent measurements

Definition of Rank

  • Rank 1 is a one-dimensional parameter
  • Rank 2 is a two-dimensional parameter
  • Rank 0 is a one-dimensional parameter describing the second dimension of a two-dimensional parameter (e.g. bin depths for moored ADCP data)

Problem Reports

No Problem Report Found in the Database


Data Access Policy

Open Data supplied by Natural Environment Research Council (NERC)

You must always use the following attribution statement to acknowledge the source of the information: "Contains data supplied by Natural Environment Research Council."


Narrative Documents

Sea-Bird Dissolved Oxygen Sensor SBE 43 and SBE 43F

The SBE 43 is a dissolved oxygen sensor designed for marine applications. It incorporates a high-performance Clark polarographic membrane with a pump that continuously plumbs water through it, preventing algal growth and the development of anoxic conditions when the sensor is taking measurements.

Two configurations are available: SBE 43 produces a voltage output and can be incorporated with any Sea-Bird CTD that accepts input from a 0-5 volt auxiliary sensor, while the SBE 43F produces a frequency output and can be integrated with an SBE 52-MP (Moored Profiler CTD) or used for OEM applications. The specifications below are common to both.

Specifications

Housing Plastic or titanium
Membrane

0.5 mil- fast response, typical for profile applications

1 mil- slower response, typical for moored applications

Depth rating

600 m (plastic) or 7000 m (titanium)

10500 m titanium housing available on request

Measurement range 120% of surface saturation
Initial accuracy 2% of saturation
Typical stability 0.5% per 1000 h

Further details can be found in the manufacturer's specification sheet.

Sea-Bird Electronics SBE 911 and SBE 917 series CTD profilers

The SBE 911 and SBE 917 series of conductivity-temperature-depth (CTD) units are used to collect hydrographic profiles, including temperature, conductivity and pressure as standard. Each profiler consists of an underwater unit and deck unit or SEARAM. Auxiliary sensors, such as fluorometers, dissolved oxygen sensors and transmissometers, and carousel water samplers are commonly added to the underwater unit.

Underwater unit

The CTD underwater unit (SBE 9 or SBE 9 plus) comprises a protective cage (usually with a carousel water sampler), including a main pressure housing containing power supplies, acquisition electronics, telemetry circuitry, and a suite of modular sensors. The original SBE 9 incorporated Sea-Bird's standard modular SBE 3 temperature sensor and SBE 4 conductivity sensor, and a Paroscientific Digiquartz pressure sensor. The conductivity cell was connected to a pump-fed plastic tubing circuit that could include auxiliary sensors. Each SBE 9 unit was custom built to individual specification. The SBE 9 was replaced in 1997 by an off-the-shelf version, termed the SBE 9 plus, that incorporated the SBE 3 plus (or SBE 3P) temperature sensor, SBE 4C conductivity sensor and a Paroscientific Digiquartz pressure sensor. Sensors could be connected to a pump-fed plastic tubing circuit or stand-alone.

Temperature, conductivity and pressure sensors

The conductivity, temperature, and pressure sensors supplied with Sea-Bird CTD systems have outputs in the form of variable frequencies, which are measured using high-speed parallel counters. The resulting count totals are converted to numeric representations of the original frequencies, which bear a direct relationship to temperature, conductivity or pressure. Sampling frequencies for these sensors are typically set at 24 Hz.

The temperature sensing element is a glass-coated thermistor bead, pressure-protected inside a stainless steel tube, while the conductivity sensing element is a cylindrical, flow-through, borosilicate glass cell with three internal platinum electrodes. Thermistor resistance or conductivity cell resistance, respectively, is the controlling element in an optimized Wien Bridge oscillator circuit, which produces a frequency output that can be converted to a temperature or conductivity reading. These sensors are available with depth ratings of 6800 m (aluminium housing) or 10500 m (titanium housing). The Paroscientific Digiquartz pressure sensor comprises a quartz crystal resonator that responds to pressure-induced stress, and temperature is measured for thermal compensation of the calculated pressure.

Additional sensors

Optional sensors for dissolved oxygen, pH, light transmission, fluorescence and others do not require the very high levels of resolution needed in the primary CTD channels, nor do these sensors generally offer variable frequency outputs. Accordingly, signals from the auxiliary sensors are acquired using a conventional voltage-input multiplexed A/D converter (optional). Some Sea-Bird CTDs use a strain gauge pressure sensor (Senso-Metrics) in which case their pressure output data is in the same form as that from the auxiliary sensors as described above.

Deck unit or SEARAM

Each underwater unit is connected to a power supply and data logging system: the SBE 11 (or SBE 11 plus) deck unit allows real-time interfacing between the deck and the underwater unit via a conductive wire, while the submersible SBE 17 (or SBE 17 plus) SEARAM plugs directly into the underwater unit and data are downloaded on recovery of the CTD. The combination of SBE 9 and SBE 17 or SBE 11 are termed SBE 917 or SBE 911, respectively, while the combinations of SBE 9 plus and SBE 17 plus or SBE 11 plus are termed SBE 917 plus or SBE 911 plus.

Specifications

Specifications for the SBE 9 plus underwater unit are listed below:

Parameter Range Initial accuracy Resolution at 24 Hz Response time
Temperature -5 to 35°C 0.001°C 0.0002°C 0.065 sec
Conductivity 0 to 7 S m-1 0.0003 S m-1 0.00004 S m-1 0.065 sec (pumped)
Pressure 0 to full scale (1400, 2000, 4200, 6800 or 10500 m) 0.015% of full scale 0.001% of full scale 0.015 sec

Further details can be found in the manufacturer's specification sheet.

Chelsea Technologies Group Aquatracka MKIII fluorometer

The Chelsea Technologies Group Aquatracka MKIII is a logarithmic response fluorometer. Filters are available to enable the instrument to measure chlorophyll, rhodamine, fluorescein and turbidity.

It uses a pulsed (5.5 Hz) xenon light source discharging along two signal paths to eliminate variations in the flashlamp intensity. The reference path measures the intensity of the light source whilst the signal path measures the intensity of the light emitted from the specimen under test. The reference signal and the emitted light signals are then applied to a ratiometric circuit. In this circuit, the ratio of returned signal to reference signal is computed and scaled logarithmically to achieve a wide dynamic range. The logarithmic conversion accuracy is maintained at better than one percent of the reading over the full output range of the instrument.

Two variants of the instrument are available, both manufactured in titanium, capable of operating in depths from shallow water down to 2000 m and 6000 m respectively. The optical characteristics of the instrument in its different detection modes are visible below:

Excitation Chlorophyll a Rhodamine Fluorescein Turbidity
Wavelength (nm) 430 500 485 440*
Bandwidth (nm) 105 70 22 80*
Emission Chlorophyll a Rhodamine Fluorescein Turbidity
Wavelength (nm) 685 590 530 440*
Bandwidth (nm) 30 45 30 80*

* The wavelengths for the turbidity filters are customer selectable but must be in the range 400 to 700 nm. The same wavelength is used in the excitation path and the emission path.

The instrument measures chlorophyll a, rhodamine and fluorescein with a concentration range of 0.01 µg l-1 to 100 µg l-1. The concentration range for turbidity is 0.01 to 100 FTU (other wavelengths are available on request).

The instrument accuracy is ± 0.02 µg l-1 (or ± 3% of the reading, whichever is greater) for chlorophyll a, rhodamine and fluorescein. The accuracy for turbidity, over a 0 - 10 FTU range, is ± 0.02 FTU (or ± 3% of the reading, whichever is greater).

Further details are available from the Aquatracka MKIII specification sheet.

Biospherical Instruments Log Quantum Cosine Irradiance Sensor QCD-905L

The QCD-905L is a submersible radiometer designed to measure irradiance over Photosynthetically Active Radiation (PAR) wavelengths (400-700 nm). It features a cosine directional response when fully immersed in water.

The sensor is a blue-enhanced high stability silicon photovoltaic detector with dielectric and absorbing glass filter assembly, and produces a logarithmic output. Normal output range is -1 to 6 volts with 1 volt per decade. Typically, the instrument outputs 5 volts for full sunlight and has a minimum output of 0.001% full sunlight, where typical noon solar irradiance is 1.5 to 2 x 1017 quanta cm-2 s-1. The instrument can be calibrated with constants for µE cm-2 s-1 or quanta cm-2 s-1.

The QCD-905L can be coupled to a fixed range data acquisition system like a CTD (Conductivity-Temperature-Depth) profiler or current meter. It has an aluminium and PET housing, and a depth rating of 7000 m.

Specifications

Wavelength 400 to 700 nm
Output range -1 to 6 V, with 1 V decade-1
Operating temperature -2 to 35°C
Depth range 0 - 7000 m

Further details can be found in the manufacturer's manual.

Tritech Digital Precision Altimeter PA200

This altimeter is a sonar ranging device that gives the height above the sea bed when mounted vertically. When mounted in any other attitude the sensor provides a subsea distance. It can be configured to operate on its own or under control from an external unit and can be supplied with simultaneous analogue and digital outputs, allowing them to interface to PC devices, data loggers, telemetry systems and multiplexers.

These instruments can be supplied with different housings, stainless steel, plastic and aluminum, which will limit the depth rating. There are three models available: the PA200-20S, PA200-10L and the PA500-6S, whose transducer options differ slightly.

Specifications

Transducer options PA200-20S P200-10L PA500-6S
Frequency (kHz) 200 200 500
Beamwidth (°) 20 Conical 10 included conical beam 6 Conical
Operating range

1 to 100 m

0.7 to 50 m

-

0.3 to 50 m

0.1 to 10 m

Common specifications are presented below

Digital resolution 1 mm
Analogue resolution 0.25% of range
Depth rating 700 , 2000, 4000 and 6800 m
Operating temperature -10 to 40°C

Further details can be found in the manufacturer's specification sheet.

WETLabs C-Star transmissometer

This instrument is designed to measure beam transmittance by submersion or with an optional flow tube for pumped applications. It can be used in profiles, moorings or as part of an underway system.

Two models are available, a 25 cm pathlength, which can be built in aluminum or co-polymer, and a 10 cm pathlength with a plastic housing. Both have an analog output, but a digital model is also available.

This instrument has been updated to provide a high resolution RS232 data output, while maintaining the same design and characteristics.

Specifications

Pathlength 10 or 25 cm
Wavelength 370, 470, 530 or 660 nm
Bandwidth

~ 20 nm for wavelengths of 470, 530 and 660 nm

~ 10 to 12 nm for a wavelength of 370 nm

Temperature error 0.02 % full scale °C-1
Temperature range 0 to 30°C
Rated depth

600 m (plastic housing)

6000 m (aluminum housing)

Further details are available in the manufacturer's specification sheet or user guide.

BODC Processing

The data were supplied to BODC in 32 MStar files and converted to the BODC internal format (QXF).

During transfer the originator's variables were mapped to unique BODC parameter codes. The following table shows the parameter mapping.

Originator's variable Units Description BODC Code Units Comments
scan - Scan number for CTD - - Not transferred - will be superseded in BODC processing
time - Time - - Not transferred - will be superseded in BODC processing
press decibars Pressure (spatial co-ordinate) exerted by the water body by profiling pressure sensor and corrected to read zero at sea level. PRESPR01 decibars -
pressure_temp degC
(ITS-90)
Pressure sensor electronics temperature - degC
(ITS-90)
Not transferred - not an environmental variable
temp degC
(ITS-90)
Temperature of the water body by CTD and NO verification against independent measurements TEMPCU01 degC
(ITS-90)
-
temp2 degC
(ITS-90)
Temperature of the water body by CTD and NO verification against independent measurements - - Secondary channel, not retained
cond mS/cm Electrical conductivity of the water body by in-situ conductivity cell and calibration against independent measurements CNCLCCI1 Siemens per metre cond divided by 10
cond2 mS/cm Electrical conductivity of the water body by CTDs - - Secondary channel, not retained
altimeter metres Height above bed in the water body AHSFZZ01 metres -
oxygen ml/l Concentration of oxygen {O2} per unit volume of the water body [dissolved phase] by Sea-Bird 43 sensor and no calibration against sample data DOXYSU01 Micromoles per litre * 44.66
fluor ug/l Concentration of chlorophyll-a {chl-a} per unit volume of the water body [particulate phase] by in-situ chlorophyll fluorometer and manufacturer's calibration applied CPHLPM01 mg/m3 ug/l=mg/m3
transmittance % Transmittance per unspecified length of the water column by transmissometer POPTPZ01 % -
par Volts Downwelling vector irradiance as photons (PAR wavelengths) in the water body by cosine-collector radiometer LVLTPU01 Volts -
psal pss-78 Practical salinity of the water body by CTD and computation using UNESCO 1983 algorithm and calibration against independent measurements PSALCC01 pss-78 -
psal2 pss-78 Practical salinity of the water body by CTD and computation using UNESCO 1983 algorithm and no calibration against independent measurements - - Secondary channel, not retained
depth metres Depth below surface of the water body by profiling pressure sensor and converted to seawater depth using UNESCO algorithm DEPHPR01 metres -
potemp degC
(ITS-90)
Potential Temperature - - Not transferred - can be calculated from pressure, salinity and temperature
potemp2 degC
(ITS-90)
Potential Temperature - - Not transferred - can be calculated from pressure, salinity and temperature

Following transfer the data were screened using BODC in-house visualisation software. Suspect data values were assigned the appropriate BODC data quality flag. Missing data values were changed to the missing data value and assigned a BODC data quality flag.

Instrument Description

CTD Unit and Auxillary Sensors

The CTD system used on cruise JR194 was the Sea-Bird 911 plus. This was mounted on a stainless steel rosette frame, equipped with 24 10-litre Niskin bottles. The CTD was fitted with the following scientific sensors:

Sensor Serial Number Last calibration date
Primary Temperature SBE-3P 4302 18th July 2007
Secondary Temperature SBE-3P 4235 20th July 2007
Primary Conductivity SBE-4C 2875 18th July 2007
Secondary Conductivity SBE-4C 2813 17th July 2007
Pressure-Digiquartz with TC 0541-75429 18th July 2007
Biospherical/Licor PAR/irradiance sensor 7235 26th July 2007
Sea-Bird SBE 43 oxygen sensor 0676 3rd June 2006
Altimeter 7742.163162 -
Chelsea Aquatracka Mk III (chlorophyll a) fluorometer 088-249 13th September 2007
Transmissometer, Chelsea/Seatech/Wetlab CStar CST-527DR 14th August 2007

Originator's Data Processing

Sampling strategy

A Conductivity-Temperature-Depth (CTD) unit was used on JR194 to produce vertical profiles of the temperature and salinity of the water column. Thirty-two stations were occupied across the Drake Passage SR1b transect, the first two of which were test dips.

Data Acquisition and Initial Processing

The CTD data were logged via the deck unit to a 1.4 GHz P4 PC, running Seasave Win32 version 5.28e (Sea-Bird Electronics Inc.). Initial processing was on the PC, applying the cell "thermal mass" correction, and then data were ported to the Unix system. Data were examined on-board; however, later, a full reprocessing of the CTD data was performed at NOCS.

The processed data, together with the raw Sea-Bird, configuration and bottle files, were supplied to BODC for banking.


Project Information

Oceans 2025 Theme 10

Oceans 2025 is a strategic marine science programme, bringing marine researchers together to increase people's knowledge of the marine environment so that they are better able to protect it for future generations.

Theme 10: Integration of Sustained Observations in the Marine Environment spans all marine domains from the sea-shore to the global ocean, providing data and knowledge on a wide range of ecosystem properties and processes (from ocean circulation to biodiversity) that are critical to understanding Earth system behaviour and identifying change. They have been developed not merely to provide long-term data sets, but to capture extreme or episodic events, and play a key role in the initialisation and validation of models. Many of these SOs will be integrated into the newly developing UK Marine Monitoring Strategy - evolving from the Defra reports Safeguarding our Seas (2002) and Charting Progress (2005), thus contributing to the underpinning knowledge for national marine stewardship. They will also contribute to the UK GOOS Strategic Plan (IACMST, 2006) and the Global Marine Assessment.

Weblink: http://www.oceans2025.org/


Oceans 2025 Theme 10, Sustained Observation Activity 6: Antarctic Circumpolar Current: Transport and Properties

Sustained Observation Activity (SO) 6 continues measurements started in 1993 of the transport of the Antarctic Circumpolar Current (ACC) in the Drake Passage, a strategic choke point for the global ocean circulation which impacts on the North Atlantic and the Meridional Overturning Circulation (MOC).

This SO contains two elements. SO 6.1. relates to the hydrographic sections and is supported by the National Oceanography Centre, Southampton (NOCS). SO 6.2. concerns bottom pressure recorders and is supported by the Proudman Oceanographic Laboratory (POL).

Aims and purpose of SO 6.1.

  • The aim is to maintain an annual hydrographic section across the ACC at Drake Passage. The standard section includes measurements of the physical properties: temperature, salinity and currents. There is the opportunity to include biogeochemical tracers, especially elements of the carbon system.

    There is need for an annual survey since the differences in transport between the annual sections are significant (~ 10% of the mean). Therefore occasional sections, say every 5 years, could only detect very large changes. Annual sections, in combination with analyses of model output, provide the possibility of extracting a long-term trend from individual measurements of a noisy signal.

Aims and purpose of SO 6.2.

  • To maintain the ocean bottom pressure and temperature time series from Drake Passage
  • To adapt the monitoring system to maximize the value of other observational programmes in the region
  • To provide better context for interpretation of the existing time series, without jeopardizing the continuity of the time series

More detailed information on this Work Package is available at pages 18 - 21 of the official Oceans 2025 Theme 10 document: Oceans 2025 Theme 10

Weblink: http://www.oceans2025.org/


Oceans 2025 - The NERC Marine Centres' Strategic Research Programme 2007-2012

Who funds the programme?

The Natural Environment Research Council (NERC) funds the Oceans 2025 programme, which was originally planned in the context of NERC's 2002-2007 strategy and later realigned to NERC's subsequent strategy (Next Generation Science for Planet Earth; NERC 2007).

Who is involved in the programme?

The Oceans 2025 programme was designed by and is to be implemented through seven leading UK marine centres. The marine centres work together in coordination and are also supported by cooperation and input from government bodies, universities and other partners. The seven marine centres are:

  • National Oceanography Centre, Southampton (NOCS)
  • Plymouth Marine Laboratory (PML)
  • Marine Biological Association (MBA)
  • Sir Alister Hardy Foundation for Marine Science (SAHFOS)
  • Proudman Oceanographic Laboratory (POL)
  • Scottish Association for Marine Science (SAMS)
  • Sea Mammal Research Unit (SMRU)

Oceans2025 provides funding to three national marine facilities, which provide services to the wider UK marine community, in addition to the Oceans 2025 community. These facilities are:

  • British Oceanographic Data Centre (BODC), hosted at POL
  • Permanent Service for Mean Sea Level (PSMSL), hosted at POL
  • Culture Collection of Algae and Protozoa (CCAP), hosted at SAMS

The NERC-run Strategic Ocean Funding Initiative (SOFI) provides additional support to the programme by funding additional research projects and studentships that closely complement the Oceans 2025 programme, primarily through universities.

What is the programme about?

Oceans 2025 sets out to address some key challenges that face the UK as a result of a changing marine environment. The research funded through the programme sets out to increase understanding of the size, nature and impacts of these changes, with the aim to:

  • improve knowledge of how the seas behave, not just now but in the future;
  • help assess what that might mean for the Earth system and for society;
  • assist in developing sustainable solutions for the management of marine resources for future generations;
  • enhance the research capabilities and facilities available for UK marine science.

In order to address these aims there are nine science themes supported by the Oceans 2025 programme:

  • Climate, circulation and sea level (Theme 1)
  • Marine biogeochemical cycles (Theme 2)
  • Shelf and coastal processes (Theme 3)
  • Biodiversity and ecosystem functioning (Theme 4)
  • Continental margins and deep ocean (Theme 5)
  • Sustainable marine resources (Theme 6)
  • Technology development (Theme 8)
  • Next generation ocean prediction (Theme 9)
  • Integration of sustained observations in the marine environment (Theme 10)

In the original programme proposal there was a theme on health and human impacts (Theme 7). The elements of this Theme have subsequently been included in Themes 3 and 9.

When is the programme active?

The programme started in April 2007 with funding for 5 years.

Brief summary of the programme fieldwork/data

Programme fieldwork and data collection are to be achieved through:

  • physical, biological and chemical parameters sampling throughout the North and South Atlantic during collaborative research cruises aboard NERC's research vessels RRS Discovery, RRS James Cook and RRS James Clark Ross;
  • the Continuous Plankton Recorder being deployed by SAHFOS in the North Atlantic and North Pacific on 'ships of opportunity';
  • physical parameters measured and relayed in near real-time by fixed moorings and ARGO floats;
  • coastal and shelf sea observatory data (Liverpool Bay Coastal Observatory (LBCO) and Western Channel Observatory (WCO)) using the RV Prince Madog and RV Quest.

The data is to be fed into models for validation and future projections. Greater detail can be found in the Theme documents.


Data Activity or Cruise Information

Cruise

Cruise Name JR20081212 (JR194, JR197)
Departure Date 2008-12-12
Arrival Date 2008-12-23
Principal Scientist(s)Graham D Quartly (National Oceanography Centre, Southampton), Miguel Angel Morales Maqueda (Proudman Oceanographic Laboratory)
Ship RRS James Clark Ross

Complete Cruise Metadata Report is available here


Fixed Station Information


No Fixed Station Information held for the Series


BODC Quality Control Flags

The following single character qualifying flags may be associated with one or more individual parameters with a data cycle:

Flag Description
Blank Unqualified
< Below detection limit
> In excess of quoted value
A Taxonomic flag for affinis (aff.)
B Beginning of CTD Down/Up Cast
C Taxonomic flag for confer (cf.)
D Thermometric depth
E End of CTD Down/Up Cast
G Non-taxonomic biological characteristic uncertainty
H Extrapolated value
I Taxonomic flag for single species (sp.)
K Improbable value - unknown quality control source
L Improbable value - originator's quality control
M Improbable value - BODC quality control
N Null value
O Improbable value - user quality control
P Trace/calm
Q Indeterminate
R Replacement value
S Estimated value
T Interpolated value
U Uncalibrated
W Control value
X Excessive difference

SeaDataNet Quality Control Flags

The following single character qualifying flags may be associated with one or more individual parameters with a data cycle:

Flag Description
0 no quality control
1 good value
2 probably good value
3 probably bad value
4 bad value
5 changed value
6 value below detection
7 value in excess
8 interpolated value
9 missing value
A value phenomenon uncertain
B nominal value
Q value below limit of quantification