Search the data

Metadata Report for BODC Series Reference Number 1812228


Metadata Summary

Data Description

Data Category CTD or STD cast
Instrument Type
NameCategories
Sea-Bird SBE 43 Dissolved Oxygen Sensor  dissolved gas sensors
Chelsea Technologies Group Aquatracka fluorometer  fluorometers
Sea-Bird SBE 911plus CTD  CTD; water temperature sensor; salinity sensor
Tritech PA-200 Altimeter  altimeters
WET Labs {Sea-Bird WETLabs} C-Star transmissometer  transmissometers
Sea-Bird SBE 3plus (SBE 3P) temperature sensor  water temperature sensor
Sea-Bird SBE 4C conductivity sensor  salinity sensor
LI-COR LI-193 Spherical Quantum PAR sensor  radiometers
Paroscientific Digiquartz depth sensors  water pressure sensors
Instrument Mounting lowered unmanned submersible
Originating Country United Kingdom
Originator Mr Hugh Venables
Originating Organization British Antarctic Survey
Processing Status banked
Online delivery of data Download available - Ocean Data View (ODV) format
Project(s) DIMES
 

Data Identifiers

Originator's Identifier CTD_JR299_002_2DB_DN_CALOX
BODC Series Reference 1812228
 

Time Co-ordinates(UT)

Start Time (yyyy-mm-dd hh:mm) 2014-03-12 19:47
End Time (yyyy-mm-dd hh:mm) -
Nominal Cycle Interval 2.0 decibars
 

Spatial Co-ordinates

Latitude 62.76520 S ( 62° 45.9' S )
Longitude 59.10950 W ( 59° 6.6' W )
Positional Uncertainty 0.0 to 0.01 n.miles
Minimum Sensor or Sampling Depth 2.97 m
Maximum Sensor or Sampling Depth 1424.16 m
Minimum Sensor or Sampling Height -
Maximum Sensor or Sampling Height -
Sea Floor Depth -
Sea Floor Depth Source -
Sensor or Sampling Distribution Variable common depth - All sensors are grouped effectively at the same depth, but this depth varies significantly during the series
Sensor or Sampling Depth Datum Instantaneous - Depth measured below water line or instantaneous water body surface
Sea Floor Depth Datum Instantaneous - Depth measured below water line or instantaneous water body surface
 

Parameters

BODC CODERankUnitsTitle
AHSFZZ011MetresHeight (spatial coordinate) relative to bed surface in the water body
CNCLCCI11Siemens per metreElectrical conductivity of the water body by in-situ conductivity cell and calibration against independent measurements
CPHLPM011Milligrams per cubic metreConcentration of chlorophyll-a {chl-a CAS 479-61-8} per unit volume of the water body [particulate >unknown phase] by in-situ chlorophyll fluorometer and manufacturer's calibration applied
DOXYSC011Micromoles per litreConcentration of oxygen {O2 CAS 7782-44-7} per unit volume of the water body [dissolved plus reactive particulate phase] by Sea-Bird SBE 43 sensor and calibration against sample data
IRRDUV011MicroEinsteins per square metre per secondDownwelling vector irradiance as photons of electromagnetic radiation (PAR wavelengths) in the water body by cosine-collector radiometer
OXYSZZ011PercentSaturation of oxygen {O2 CAS 7782-44-7} in the water body [dissolved plus reactive particulate phase]
POPTZZ011PercentTransmittance (unspecified wavelength) per unspecified length of the water body by transmissometer
POTMCV011Degrees CelsiusPotential temperature of the water body by computation using UNESCO 1983 algorithm
PRESPR011DecibarsPressure (spatial coordinate) exerted by the water body by profiling pressure sensor and correction to read zero at sea level
PSALST011DimensionlessPractical salinity of the water body by CTD and computation using UNESCO 1983 algorithm
SIGTPR011Kilograms per cubic metreSigma-theta of the water body by CTD and computation from salinity and potential temperature using UNESCO algorithm
TEMPST011Degrees CelsiusTemperature of the water body by CTD or STD
TOKGPR011Litres per kilogramConversion factor (volume to mass) for the water body by CTD and computation of density (in-situ potential temperature surface pressure) reciprocal from pressure, temperature and salinity

Definition of Rank

  • Rank 1 is a one-dimensional parameter
  • Rank 2 is a two-dimensional parameter
  • Rank 0 is a one-dimensional parameter describing the second dimension of a two-dimensional parameter (e.g. bin depths for moored ADCP data)

Problem Reports

No Problem Report Found in the Database


Data Access Policy

Open Data

These data have no specific confidentiality restrictions for users. However, users must acknowledge data sources as it is not ethical to publish data without proper attribution. Any publication or other output resulting from usage of the data should include an acknowledgment.

If the Information Provider does not provide a specific attribution statement, or if you are using Information from several Information Providers and multiple attributions are not practical in your product or application, you may consider using the following:

"Contains public sector information licensed under the Open Government Licence v1.0."


Narrative Documents

LI-COR LI-193 Spherical Quantum Sensor

The LI-193 Underwater Spherical Quantum Sensor measures irradiance over the photosynthetically active radiation (PAR) spectral range (400-700 nm). The instrument measures photon flux from all directions, which is often termed photosynthetic photon flux fluence rate (PPFFR) or quantum scalar irradiance. This measurement is particularly important when studying phytoplankton, which utilize radiation from all directions for photosynthesis.

The LI-193 can operate down to 350 m and is fitted with a blue-enhanced silicon photovoltaic detector. Two versions of the instrument are available (193SA and 193SB). These use different connector types for the underwater cable, but the sensors operate in the same manner.

Sensor Specifications

Absolute Calibration ± 5 % in air traceable to NBS.
Sensitivity Typically 7 µA per 1000 µmol s-1 m-2
Linearity Maximum deviation of 1 % up to 10,000 µmol s-1 m-2.
Stability < ± 2 % change over a 1 year period.
Response Time 10 µs.
Temperature Dependence ± 0.15 % per °C maximum.
Angular Response < ± 4% error up to ± 90° from normal axis.
Azimuth < ± 3 % error over 360 ° at 90 ° from normal axis.
Detector High stability silicon photovoltaic detector (blue enhanced).
Sensor Housing Corrosion resistant metal with acrylic diffuser for both saltwater and freshwater applications with an injection moulded, impact resistant, acrylic diffuser. Units have been tested to 500 psi (3400 kPa) (350 meters).
Weights 142 g
Cable Requires 2222UWB Underwater Cable

Further information can be found in the manufacturer's specification sheet and instruction manual.

Sea-Bird Dissolved Oxygen Sensor SBE 43 and SBE 43F

The SBE 43 is a dissolved oxygen sensor designed for marine applications. It incorporates a high-performance Clark polarographic membrane with a pump that continuously plumbs water through it, preventing algal growth and the development of anoxic conditions when the sensor is taking measurements.

Two configurations are available: SBE 43 produces a voltage output and can be incorporated with any Sea-Bird CTD that accepts input from a 0-5 volt auxiliary sensor, while the SBE 43F produces a frequency output and can be integrated with an SBE 52-MP (Moored Profiler CTD) or used for OEM applications. The specifications below are common to both.

Specifications

Housing Plastic or titanium
Membrane

0.5 mil- fast response, typical for profile applications

1 mil- slower response, typical for moored applications

Depth rating

600 m (plastic) or 7000 m (titanium)

10500 m titanium housing available on request

Measurement range 120% of surface saturation
Initial accuracy 2% of saturation
Typical stability 0.5% per 1000 h

Further details can be found in the manufacturer's specification sheet.

Instrument Description

CTD Unit and Auxillary Sensors

A Sea-Bird 11plus CTD system was used on cruise JR299 Leg 1 (JR20140309) and Leg 2 (JR20140331). This was mounted on a 24-way stainless steel rosette frame, equipped with 24 12 litre Niskin bottles.

The CTD was fitted with the following scientific sensors:

Sensor Serial Number Last calibration date Comments
Primary Temperature SBE 3Plus 4472 12th March 2013 -
Secondary Temperature SBE 3Plus 2366 30th August 2012 -
Primary Conductivity SBE 4C 2875 13th March 2013 -
Secondary Conductivity SBE 4C 2289 21stAugust 2012 -
Tritech PA200 Altimeter 244740 16th May 2012 -
Sea-Bird SBE 43 dissolved oxygen sensor 2290 31st March 2012 CTD001-CTD037
Sea-Bird SBE 43 dissolved oxygen sensor 0676 28th August 2012 CTD038-CTD118
Transmissometer WET Labs C-Star CST-846DR 13th March 2013 -
Chelsea Aquatracka MKIII fluorometer 88216 19th February 2013 -
Biospherical LI-COR Chelsea PAR sensor 7235 24th April 2013 -
Paroscientific Digiquartz with TC depth sensor 89973 22nd August 2012 -

Sea-Bird Electronics SBE 911 and SBE 917 series CTD profilers

The SBE 911 and SBE 917 series of conductivity-temperature-depth (CTD) units are used to collect hydrographic profiles, including temperature, conductivity and pressure as standard. Each profiler consists of an underwater unit and deck unit or SEARAM. Auxiliary sensors, such as fluorometers, dissolved oxygen sensors and transmissometers, and carousel water samplers are commonly added to the underwater unit.

Underwater unit

The CTD underwater unit (SBE 9 or SBE 9 plus) comprises a protective cage (usually with a carousel water sampler), including a main pressure housing containing power supplies, acquisition electronics, telemetry circuitry, and a suite of modular sensors. The original SBE 9 incorporated Sea-Bird's standard modular SBE 3 temperature sensor and SBE 4 conductivity sensor, and a Paroscientific Digiquartz pressure sensor. The conductivity cell was connected to a pump-fed plastic tubing circuit that could include auxiliary sensors. Each SBE 9 unit was custom built to individual specification. The SBE 9 was replaced in 1997 by an off-the-shelf version, termed the SBE 9 plus, that incorporated the SBE 3 plus (or SBE 3P) temperature sensor, SBE 4C conductivity sensor and a Paroscientific Digiquartz pressure sensor. Sensors could be connected to a pump-fed plastic tubing circuit or stand-alone.

Temperature, conductivity and pressure sensors

The conductivity, temperature, and pressure sensors supplied with Sea-Bird CTD systems have outputs in the form of variable frequencies, which are measured using high-speed parallel counters. The resulting count totals are converted to numeric representations of the original frequencies, which bear a direct relationship to temperature, conductivity or pressure. Sampling frequencies for these sensors are typically set at 24 Hz.

The temperature sensing element is a glass-coated thermistor bead, pressure-protected inside a stainless steel tube, while the conductivity sensing element is a cylindrical, flow-through, borosilicate glass cell with three internal platinum electrodes. Thermistor resistance or conductivity cell resistance, respectively, is the controlling element in an optimized Wien Bridge oscillator circuit, which produces a frequency output that can be converted to a temperature or conductivity reading. These sensors are available with depth ratings of 6800 m (aluminium housing) or 10500 m (titanium housing). The Paroscientific Digiquartz pressure sensor comprises a quartz crystal resonator that responds to pressure-induced stress, and temperature is measured for thermal compensation of the calculated pressure.

Additional sensors

Optional sensors for dissolved oxygen, pH, light transmission, fluorescence and others do not require the very high levels of resolution needed in the primary CTD channels, nor do these sensors generally offer variable frequency outputs. Accordingly, signals from the auxiliary sensors are acquired using a conventional voltage-input multiplexed A/D converter (optional). Some Sea-Bird CTDs use a strain gauge pressure sensor (Senso-Metrics) in which case their pressure output data is in the same form as that from the auxiliary sensors as described above.

Deck unit or SEARAM

Each underwater unit is connected to a power supply and data logging system: the SBE 11 (or SBE 11 plus) deck unit allows real-time interfacing between the deck and the underwater unit via a conductive wire, while the submersible SBE 17 (or SBE 17 plus) SEARAM plugs directly into the underwater unit and data are downloaded on recovery of the CTD. The combination of SBE 9 and SBE 17 or SBE 11 are termed SBE 917 or SBE 911, respectively, while the combinations of SBE 9 plus and SBE 17 plus or SBE 11 plus are termed SBE 917 plus or SBE 911 plus.

Specifications

Specifications for the SBE 9 plus underwater unit are listed below:

Parameter Range Initial accuracy Resolution at 24 Hz Response time
Temperature -5 to 35°C 0.001°C 0.0002°C 0.065 sec
Conductivity 0 to 7 S m-1 0.0003 S m-1 0.00004 S m-1 0.065 sec (pumped)
Pressure 0 to full scale (1400, 2000, 4200, 6800 or 10500 m) 0.015% of full scale 0.001% of full scale 0.015 sec

Further details can be found in the manufacturer's specification sheet.

Aquatracka fluorometer

The Chelsea Instruments Aquatracka is a logarithmic response fluorometer. It uses a pulsed (5.5 Hz) xenon light source discharging between 320 and 800 nm through a blue filter with a peak transmission of 420 nm and a bandwidth at half maximum of 100 nm. A red filter with sharp cut off, 10% transmission at 664 nm and 678 nm, is used to pass chlorophyll-a fluorescence to the sample photodiode.

The instrument may be deployed either in a through-flow tank, on a CTD frame or moored with a data logging package.

Further details can be found in the manufacturer's specification sheet.

Tritech Digital Precision Altimeter PA200

This altimeter is a sonar ranging device that gives the height above the sea bed when mounted vertically. When mounted in any other attitude the sensor provides a subsea distance. It can be configured to operate on its own or under control from an external unit and can be supplied with simultaneous analogue and digital outputs, allowing them to interface to PC devices, data loggers, telemetry systems and multiplexers.

These instruments can be supplied with different housings, stainless steel, plastic and aluminum, which will limit the depth rating. There are three models available: the PA200-20S, PA200-10L and the PA500-6S, whose transducer options differ slightly.

Specifications

Transducer options PA200-20S P200-10L PA500-6S
Frequency (kHz) 200 200 500
Beamwidth (°) 20 Conical 10 included conical beam 6 Conical
Operating range

1 to 100 m

0.7 to 50 m

-

0.3 to 50 m

0.1 to 10 m

Common specifications are presented below

Digital resolution 1 mm
Analogue resolution 0.25% of range
Depth rating 700 , 2000, 4000 and 6800 m
Operating temperature -10 to 40°C

Further details can be found in the manufacturer's specification sheet.

WETLabs C-Star transmissometer

This instrument is designed to measure beam transmittance by submersion or with an optional flow tube for pumped applications. It can be used in profiles, moorings or as part of an underway system.

Two models are available, a 25 cm pathlength, which can be built in aluminum or co-polymer, and a 10 cm pathlength with a plastic housing. Both have an analog output, but a digital model is also available.

This instrument has been updated to provide a high resolution RS232 data output, while maintaining the same design and characteristics.

Specifications

Pathlength 10 or 25 cm
Wavelength 370, 470, 530 or 660 nm
Bandwidth

~ 20 nm for wavelengths of 470, 530 and 660 nm

~ 10 to 12 nm for a wavelength of 370 nm

Temperature error 0.02 % full scale °C-1
Temperature range 0 to 30°C
Rated depth

600 m (plastic housing)

6000 m (aluminum housing)

Further details are available in the manufacturer's specification sheet or user guide.

BODC Processing

The CTD data were supplied to BODC as 118 MStar files and converted to the BODC internal format.

During transfer the originator's variables were mapped to unique BODC parameter codes. The following table shows the parameter mapping.

Originator's variable Units BODC Code Units Comments
press decibars PRESPR01 decibars -
pressure_temp degC - - Not transferred - not an environmental variable
altimeter m AHSFZZ01 m -
temp degC TEMPST01 degC -
temp1 degC - - Secondary channel, not retained
temp2 degC - - Secondary channel, not retained
cond mS/cm CNCLCCI1 S/m Converted (/10)
cond1 mS/cm - - Secondary channel, not retained
cond2 mS/cm - - Secondary channel, not retained
oxygen µmol/kg DOXYSC01 µmol/l Conversion by BODC to µmol/l using TOKGPR01
fluor µg/l CPHLPM01 mg/m3 µg/l=mg/m3
transmittance % POPTZZ01 % -
PAR number IRRDUV01 µE/m2/s -
psal pss-78 PSALST01 pss-78 Calculated from calibrated conductivity measurements by the originator
sigma0 kg/m3 - - -
sigma2 kg/m3 - - -
sigma4 kg/m3 - - -
- - OXYSZZ01 % Derived by BODC using DOXYSC01, TEMPST01 and PSALST01
- - POTMCV01 °C Derived by BODC using TEMPST01, PSALST01 and PRESPR01.
- - SIGTPR01 kg/m3 Derived by BODC using POTMCV01, PSALST01 and PRESPR01
- - TOKGPR01 l/kg1 Derived by BODC using SIGTPR01

Following transfer, the data were screened using BODC in-house visualisation software. Suspect data values were assigned the appropriate BODC data quality flag. Missing data values, where present, were changed to the missing data value and assigned a BODC data quality flag. During screening, oxygen values were found to be missing for CTD casts 37-40 in the originator's files.

Cruise JR299 (JR20140309 and JR20140331) Originator's CTD Data Processing

Sampling Strategy

A total of 119 CTD casts were performed during cruise JR299 (JR20140309 and JR20140331). However, CTD 21 was not retained due to a malfuntion of the equipment. Continuous recordings of conductivity, temperature, depth, fluorescence, PAR, and transmittance were taken for each CTD cast. In addition, Lowered Acoustic Doppler Current Profilers (LADCPs) were attached to the CTD frame on all 119 CTD casts. JR299 Leg 1 (JR20140309) contains CTD casts from 08/03/2014-30/03/2014 and Leg 2 (JR20140331) contains CTD casts from 31/03/2014-29/04/2014.

Data Processing

Raw CTD data were transferred from the Sea-Bird deck unit via Sea-Bird software. The binary files were converted using Sea-Bird processing software (Seasave, Version 7.22.3) to a collection of ASCII (.cnv) files. Then MEXEC programs were run to process the data which included reducing the frequency of the data from 24Hz to 1Hz, calibrating the data, calculating derived variables, and extracting information from the bottom of the cast identified by the maximum pressure. The temperature and salinity sensors were calibrated post-cruise against independent discrete samples taken from water bottles on the CTD. Details of the MEXEC programs used and further details of the processing performed can be found in the cruise report.


Project Information

Diapycnal and Isopycnal Mixing Experiment in the Southern Ocean (DIMES) project document

DIMES is a US/UK field program aimed at measuring diapycnal and isopycnal mixing in the Southern Ocean, along the tilting isopycnals of the Antarctic Circumpolar Current.

The Meridional Overturning Circulation (MOC) of the ocean is a critical regulator of the Earth's climate processes. Climate models are highly sensitive to the representation of mixing processes in the southern limb of the MOC, within the Southern Ocean, although the lack of extensive in situ observations of Southern Ocean mixing processes has made evaluation of mixing somewhat difficult. Theories and models of the Southern Ocean circulation have been built on the premise of adiabatic flow in the ocean interior, with diabatic processes confined to the upper-ocean mixed layer. Interior diapycnal mixing has often been assumed to be small, but a few recent studies have suggested that diapycnal mixing might be large in some locations, particularly over rough bathymetry. Depending on its extent, this interior diapycnal mixing could significantly affect the overall energetics and property balances for the Southern Ocean and in turn for the global ocean. The goals of DIMES are to obtain measurements that will help us quantify both along-isopycnal eddy-driven mixing and cross-isopycnal interior mixing.

DIMES includes tracer release, isopycnal following RAFOS floats, microstructure measurements, shearmeter floats, EM-APEX floats, a mooring array in Drake Passage, hydrographic observations, inverse modeling, and analysis of altimetry and numerical model output.

DIMES is sponsored by the National Science Foundation (U.S.), Natural Environment Research Council (U.K) and British Antarctic Survey (U.K.)

For more information please see the official project website at DIMES


Data Activity or Cruise Information

Cruise

Cruise Name JR20140308 (JR293 Leg1, JR299 Leg1, UKD-5)
Departure Date 2014-03-08
Arrival Date 2014-03-29
Principal Scientist(s)Andrew J S Meijers (British Antarctic Survey)
Ship RRS James Clark Ross

Complete Cruise Metadata Report is available here


Fixed Station Information


No Fixed Station Information held for the Series


BODC Quality Control Flags

The following single character qualifying flags may be associated with one or more individual parameters with a data cycle:

Flag Description
Blank Unqualified
< Below detection limit
> In excess of quoted value
A Taxonomic flag for affinis (aff.)
B Beginning of CTD Down/Up Cast
C Taxonomic flag for confer (cf.)
D Thermometric depth
E End of CTD Down/Up Cast
G Non-taxonomic biological characteristic uncertainty
H Extrapolated value
I Taxonomic flag for single species (sp.)
K Improbable value - unknown quality control source
L Improbable value - originator's quality control
M Improbable value - BODC quality control
N Null value
O Improbable value - user quality control
P Trace/calm
Q Indeterminate
R Replacement value
S Estimated value
T Interpolated value
U Uncalibrated
W Control value
X Excessive difference

SeaDataNet Quality Control Flags

The following single character qualifying flags may be associated with one or more individual parameters with a data cycle:

Flag Description
0 no quality control
1 good value
2 probably good value
3 probably bad value
4 bad value
5 changed value
6 value below detection
7 value in excess
8 interpolated value
9 missing value
A value phenomenon uncertain
B nominal value
Q value below limit of quantification