Metadata Report for BODC Series Reference Number 902004


Metadata Summary

Data Description

Data Category CTD or STD cast
Instrument Type
NameCategories
Sea-Bird SBE 19 SEACAT CTD  CTD; water temperature sensor; salinity sensor
Seapoint chlorophyll fluorometer  fluorometers
WET Labs {Sea-Bird WETLabs} C-Star transmissometer  transmissometers
Instrument Mounting research vessel
Originating Country United Kingdom
Originator Ms Sarah Hughes
Originating Organization Fisheries Research Services Aberdeen Marine Laboratory (now Marine Scotland Aberdeen Marine Laboratory)
Processing Status banked
Online delivery of data Download available - Ocean Data View (ODV) format
Project(s) -
 

Data Identifiers

Originator's Identifier SC12/07/282
BODC Series Reference 902004
 

Time Co-ordinates(UT)

Start Time (yyyy-mm-dd hh:mm) 2007-08-11 10:54
End Time (yyyy-mm-dd hh:mm) -
Nominal Cycle Interval 1.0 decibars
 

Spatial Co-ordinates

Latitude 56.29767 N ( 56° 17.9' N )
Longitude 1.02033 W ( 1° 1.2' W )
Positional Uncertainty 0.05 to 0.1 n.miles
Minimum Sensor or Sampling Depth 0.99 m
Maximum Sensor or Sampling Depth 66.38 m
Minimum Sensor or Sampling Height 4.62 m
Maximum Sensor or Sampling Height 70.01 m
Sea Floor Depth 71.0 m
Sea Floor Depth Source -
Sensor or Sampling Distribution Variable common depth - All sensors are grouped effectively at the same depth, but this depth varies significantly during the series
Sensor or Sampling Depth Datum Instantaneous - Depth measured below water line or instantaneous water body surface
Sea Floor Depth Datum Instantaneous - Depth measured below water line or instantaneous water body surface
 

Parameters

BODC CODERankUnitsTitle
ACYCAA011DimensionlessSequence number
CNCLCCI11Siemens per metreElectrical conductivity of the water body by in-situ conductivity cell and calibration against independent measurements
CPHLPM011Milligrams per cubic metreConcentration of chlorophyll-a {chl-a CAS 479-61-8} per unit volume of the water body [particulate >unknown phase] by in-situ chlorophyll fluorometer and manufacturer's calibration applied
PRESPR011DecibarsPressure (spatial coordinate) exerted by the water body by profiling pressure sensor and correction to read zero at sea level
PSALST011DimensionlessPractical salinity of the water body by CTD and computation using UNESCO 1983 algorithm
TEMPST011Degrees CelsiusTemperature of the water body by CTD or STD

Definition of Rank

  • Rank 1 is a one-dimensional parameter
  • Rank 2 is a two-dimensional parameter
  • Rank 0 is a one-dimensional parameter describing the second dimension of a two-dimensional parameter (e.g. bin depths for moored ADCP data)

Problem Reports

No Problem Report Found in the Database

Data quality report

No transmissometer used during this CTD profile.


Data Access Policy

Public domain data

These data have no specific confidentiality restrictions for users. However, users must acknowledge data sources as it is not ethical to publish data without proper attribution. Any publication or other output resulting from usage of the data should include an acknowledgment.

The recommended acknowledgment is

"This study uses data from the data source/organisation/programme, provided by the British Oceanographic Data Centre and funded by the funding body."


Narrative Documents

Instrument Description

CTD Unit and Auxiliary Sensors

A Sea-Bird Electronics SBE 19 SEACAT CTD unit with an SBE 32 carousel water sampler was used. The CTD unit included the following sensors.

Sensor Manufacturer Model Serial number Calibration date
Pressure Sea-Bird   2304 2007-04-30
Temperature Sea-Bird   2304 2007-05-04
Conductivity Sea-Bird   2304 2007-05-04
Fluorometer Seapoint Chlorophyll Fluorometer 2048 2006-02-23
Transmissometer Wet Labs C-star CST-504PR 2006-01-19

Sea-Bird SBE 19 and SBE 19plus SEACAT Profiler CTDs

The SBE 19 SEACAT Profiler is a self-contained, battery powered, pumped CTD system designed to measure conductivity, temperature, and pressure in marine or fresh water environments to depths of 10,500 meters. It was replaced by the SBE 19plus model in 2001. An updated version of this instrument is the SBE 19plus V2, which incorporates an electronics upgrade and additional features, with six differentially amplified A/D input channels, one RS-232 data input channel, and 64 MB FLASH memory.

The standard CTD unit comes with a plastic housing (rated to 600 m), although this can be replaced by titanium housing for depths up to 7000 m. It is typically used for CTD profiling although a conversion kit is available for mooring deployments. The CTD can also be attached to an SBE 36 CTD Deck Unit and Power/Data Interface Module (PDIM) for real-time operation on single-core armored cable up to 10,000 m.

Specifications

Parameter SBE 19 SBE 19plus
Temperature

Range: -5 to +35 °C

Accuracy: 0.01 °C

Resolution: 0.001 °C

Calibration: +1 to +32 °C*

Range: -5 to +35 °C

Accuracy: 0.005 °C

Resolution: 0.0001 °C

Calibration: +1 to +32 °C*

Conductivity

Range: 0 to 7 S m-1 (0 to 70 mmho cm-1)

Accuracy: 0.001 S m-1

Resolution: 0.0001 S m-1

Calibration: 0 to 7 S m-1. Physical calibration over the range 1.4 - 6 S m-1*

Range: 0 to 9 Sm-1

Accuracy: 0.0005

Resolution: 0.00005 (most oceanic waters, resolves 0.4 ppm in salinity); 0.00007 (high salinity waters, resolves 0.4 ppm in salinity); 0.00001 (fresh waters, resolves 0.1 ppm in salinity)

Calibration: 0 to 9 S m-1. Physical calibration over the range 1.4 - 6 S m-1*

Strain gauge pressure sensor

Range: 0 to100, 150, 300, 500, 1000, 1500, 3000, 5000, 10000 or 15000 psia

Accuracy: 0.25% of full scale range (100 - 1500 psia); 0.15% of full scale range (3000 - 15000 psia)

Resolution: 0.015% of full scale

Calibration: 0 to full scale in 20% steps

Range: 0 to 20, 100, 350, 1000, 2000, 3500 or 7000 m

Accuracy: 0.1% of full scale range

Resolution: 0.002% of full scale range

Calibration: ambient pressure to full scale range in 5 steps

*Measurements outside this range may be at slightly reduced accuracy due to extrapolation errors.

Options and accessories

Additional sensors can be attached to the CTD, including:

  • high accuracy Paroscientific Digiquartz pressure sensor (depth range 0 to 20, 60, 130, 200, 270, 680, 1400, 2000, 4200, 7000 or 10500 m; accuracy 0.02% of full scale; resolution 0.0025% of full scale)
  • Dissolved Oxygen (SBE 43 DO Sensor)
  • pH* (SBE 18 pH Sensor or SBE 27 pH/ORP Sensor)
  • fluorescence
  • radiance (PAR)
  • light transmission
  • optical backscatter (turbidity)

The standard SBE 5M pump may be replaced by an SBE 5P (plastic housing) or 5T (titanium housing) pump for use with dissolved oxygen and/or other pumped sensors. Further details can be found in the manufacturer's SBE 19plus V2 instrument specification or theSBE 19 andSBE 19 plus user guides.

Seapoint Chlorophyll Fluorometer

The Seapoint Chlorophyll Fluorometer (SCF) is a low power instrument for in situ measurements of chlorophyll a. The SCF uses modulated blue LED lamps and a blue excitation filter to excite chlorophyll a. The fluorescent light emitted by the chlorophyll a passes through a red emission filter and is detected by a silicon photodiode. The low level signal is then processed using synchronous demodulation circuitry which generates an output voltage proportional to chlorophyll a concentration. The SCF may be operated with or without a pump.

Sensor specifications, current at August 2006, are given in the table below. More information can be found at the manufacturer's web site.

Sensor Specifications

Power requirements 8 - 20 VDC, 15 mA avg., 27 mA pk.
Output 0 - 5.0 VDC
Output Time Constant 0.1 sec.
Power-up transient period < 1 sec.
Excitation Wavelength 470 nm CWL, 30 nm FWHM
Emission Wavelength 685 nm CWL, 30 nm FWHM
Sensing Volume 340 mm3
Minimum Detectable Level 0.02 µg l-1

  Gain Sensitivity, V µg-1 l-1 Range, µg l-1
Sensitivity/Range 30x
10x
3x
1x
1.0
0.33
0.1
0.033
5
15
50
150

WETLabs C-Star transmissometer

This instrument is designed to measure beam transmittance by submersion or with an optional flow tube for pumped applications. It can be used in profiles, moorings or as part of an underway system.

Two models are available, a 25 cm pathlength, which can be built in aluminum or co-polymer, and a 10 cm pathlength with a plastic housing. Both have an analog output, but a digital model is also available.

This instrument has been updated to provide a high resolution RS232 data output, while maintaining the same design and characteristics.

Specifications

Pathlength 10 or 25 cm
Wavelength 370, 470, 530 or 660 nm
Bandwidth

~ 20 nm for wavelengths of 470, 530 and 660 nm

~ 10 to 12 nm for a wavelength of 370 nm

Temperature error 0.02 % full scale °C-1
Temperature range 0 to 30°C
Rated depth

600 m (plastic housing)

6000 m (aluminum housing)

Further details are available in the manufacturer's specification sheet or user guide.

BODC Processing

Data were received by BODC in one ASCII format file that was subsequently split into 86 separate files, one for each CTD profile. The series were reformatted to the internal QXF format using BODC transfer function 340. Sample calibrations were applied to the conductivity data. The following table details the mapping of variables to BODC parameter codes.

Original parameter name Original Units Description BODC Parameter Code BODC Units Comments
Pressure Decibars Pressure exerted by the water column PRESPR01 Decibars  
Temperature °C Temperature of the water column TEMPST01 °C  
Conductivity mS cm-1 Electrical conductivity of the water column calibrated against independent measurements CNCLCCI1 S m-1 Conversion by transfer (mS cm-1 x 0.1). Sample calibrations applied by transfer
Salinity   Salinity of the water column PSALST01 Dimensionless Derived by transfer using UNESCO 1983 algorithm
Fluorescence µg l-1 In-situ fluorescence CPHLPM01 mg m-3  
Beam attenuation m-1 Attenuance of the water column ATTNMR01 m-1  

Following transfer to QXF, the data were screened using BODC's in-house visualisation software, EDSERPLO. Any data considered as suspect were flagged 'M'. Flags from the originator marking suspect data were retained during transfer and flagged 'L'.

Originator's Data Processing

Sampling Strategy

A total of 86 CTD casts were performed on FRV Scotia cruise 1207S (10 August 2007 - 01 September 2007) around the North Sea and Shetland. The data were collected between 04:00 and 20:00 hours during the period 10 August 2007 - 30 August 2007.

Rosette bottles were fired in order to obtain independent salinity measurements. The sample data were used to derive calibrations for the conductivity profiles collected by the CTD.

Data Processing

The raw CTD data files were processed through the SeaBird Electronics SeaSoft data processing software following standard procedures. The originators used in-house interactive visual display editing software to edit out individual spikes in the primary temperature and conductivity channels. An ASCII file was generated for each CTD cast and all files from a cruise were concatenated into one ASCII file which was submitted to BODC.

Field Calibrations

Independent salinity samples, obtained from the CTD rosette and spread throughout the cruise, were used to calibrate the CTD conductivity data. Outlying points were discarded, and between 48 and 72 data points were used to derive the calibrations. The sample analyses yielded a straight line conductivity calibration of the form y = mx + c, where m = 1.001716 and c = -0.075157.

The uncalibrated data and calibrations were submitted to BODC, who applied the appropriate corrections.


Project Information


No Project Information held for the Series

Data Activity or Cruise Information

Cruise

Cruise Name 1207S
Departure Date 2007-08-10
Arrival Date 2007-09-01
Principal Scientist(s)Kenneth A Coull (Fisheries Research Services Aberdeen Marine Laboratory)
Ship FRV Scotia

Complete Cruise Metadata Report is available here


Fixed Station Information


No Fixed Station Information held for the Series


BODC Quality Control Flags

The following single character qualifying flags may be associated with one or more individual parameters with a data cycle:

Flag Description
Blank Unqualified
< Below detection limit
> In excess of quoted value
A Taxonomic flag for affinis (aff.)
B Beginning of CTD Down/Up Cast
C Taxonomic flag for confer (cf.)
D Thermometric depth
E End of CTD Down/Up Cast
G Non-taxonomic biological characteristic uncertainty
H Extrapolated value
I Taxonomic flag for single species (sp.)
K Improbable value - unknown quality control source
L Improbable value - originator's quality control
M Improbable value - BODC quality control
N Null value
O Improbable value - user quality control
P Trace/calm
Q Indeterminate
R Replacement value
S Estimated value
T Interpolated value
U Uncalibrated
W Control value
X Excessive difference

SeaDataNet Quality Control Flags

The following single character qualifying flags may be associated with one or more individual parameters with a data cycle:

Flag Description
0 no quality control
1 good value
2 probably good value
3 probably bad value
4 bad value
5 changed value
6 value below detection
7 value in excess
8 interpolated value
9 missing value
A value phenomenon uncertain
B nominal value
Q value below limit of quantification